中文 |

Line-scan Raman micro-spectroscopy Provides Rapid Method for Micro and Nanoplastics Detection

Author: WU Qingyi |

Microplastics, as plastics particles smaller than 5mm in size, have caused an environmental pollution issue that cannot be ignored by our society. Raman spectroscopy technology, with its non- -contact, non-destructive, and chemical-specific characteristics, has been widely applied in the field of microplastics detection. However, conventional point confocal Raman techniques are limited to single-point detection, impeding their detection speed.

In a study published in Talanta, a research group led by Prof. LI Bei from the Changchun Institute of Optics, Fine Mechanics and Physics (CIOMP) of the Chinese Academy of Sciences (CAS), in collaboration with Prof. Wolfgang Langbein from Cardiff University, proposed a novel line-scan Raman micro-spectroscopy technique for rapid identification of micro and nanoplastics.

Based on the fundamental principles of confocal Raman spectroscopy, the focused excitation spot transforms from a convergent point into a convergent line with diffraction-limited width. The optical setup employs a conjugate imaging design. In the two-dimensional image recorded by the charge-coupled device (CCD), the vertical dimension maps the vertical dimension of the sample along the excitation line, while the spectrum is dispersed along the horizontal dimension. In this way, a single acquisition provides the spectra for all spatial positions along the excitation line.

Researchers developed a confocal line-scan Raman micro-spectroscopy system, established a preprocessing workflow for line-scan Raman spectral data, and applied the factorization into susceptibilities and concentrations (FSC3) algorithm to obtain Raman hyperspectral images. The team employed a concave cylindrical lens to generate the excitation line and improved the uniformity of energy distribution using a Powell lens.

Plastic beads of various sizes were used for size and composition identification. The detection of beads with a diameter of 200 nm, which is smaller than the diffraction limit, was realized, demonstrating the exceptional sensitivity of the line-scan Raman spectroscopy system.

Furthermore, four types of plastic powder samples were used for a large-scale area of 1.2mm in length and 40μm in height measurement. Impressively, the imaging time is 20 minutes to obtain a 240000-pixel Raman image. Compared with point confocal Raman imaging, the line-scan confocal Raman technology increases the imaging speed by 2 orders of magnitude.

Line-scan Raman micro-spectroscopy offers non-destructive analysis with high sensitivity and high-throughput. By employing appropriate sampling techniques such as filtration or sedimentation, environmental samples from various sources, including water, soil, and air, are accessible.

Contact

LI Bei

Changchun Institute of Optics, Fine Mechanics and Physics

E-mail:




       Copyright @ 吉ICP备06002510号 2007 CIOMP130033